Home    HUMAN    Healfo SARS-CoV-2 IgG/IgM Cassette Rapid Test
IgGIgMCassetteRapidTest225×125×70

Healfo SARS-CoV-2 IgG/IgM Cassette Rapid Test

INTENDED USE

The Healfo SARS-CoV-2 IgG/IgM Cassette Rapid Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of IgM and IgG antibodies to SARS-CoV-2 in serum, plasma (EDTA, citrate) or venipuncture whole blood specimens from patients suspected of COVID-19 infection by a healthcare provider. The Healfo SARS-CoV-2 IgG/IgM Cassette Rapid Test is an aid in the diagnosis of patients with suspected SARS-CoV-2 infection in conjunction with clinical presentation and the results of other laboratory tests. Results from the Healfo SARS-CoV-2 IgG/IgM Cassette Rapid Test should not be used as the sole basis for diagnosis.

For prescription use only. For in vitro diagnostic use only. For emergency use authorization use only.

TEST PRINCIPLE

The Healfo SARS-CoV-2 IgG/IgM Cassette Rapid Test is a lateral flow chromatographic immunoassay.

img2

The SARS-CoV-2 recombinant antigens labeled by colloidal gold (SARS-CoV-2 conjugates) in burgundy colored conjugate pad can bind to the antibodies against the SARS-CoV-2 virus, such that if the anti-SARS-CoV-2 virus IgG or IgM or both present in the specimen, the IgG or IgM can be binded with the conjugates to form the IgG- SARS-CoV-2 conjugates or IgM-SARS-CoV-2 conjugates immunocomplex.

Migrating through the nitrocellulose matrix, the immunocomplex of IgM-(SARS-CoV-2) conjugates will then captured by the anti-human IgM line, forming a burgundy colored GMLine. The result is IgM positive or reactive. Information regarding the immune response to SARS-CoV-2 is limited and still evolving.

And the immunocomplex of IgG- (SARS-CoV-2) conjugates will then be captured by the anti-human IgG line, forming a burgundy-colored G Line. The result is IgG positive or reactive, consistent with a recent or previous infection.

This test contains a built-in control feature, the C Line. The goat anti-rabbit IgG bound in the C line, can bind with the rabbit IgG-gold conjugates which from the conjugate pad. The C Line develops after addition of the specimen and sample diluent. If the C Line does not develop, the assay is invalid regardless of color development of the G or M Lines as indicated below. Repeat the assay with a new device. 

MATERIAL

Material Provided:

There are three kit sizes. The kit component configurations are provided below: 

Components

EA/test

Total

Kit Size (#of Tests)

1

25

Test Cassette (#)

1

25

Sample Diluent (# of Bottles)

1

1

Transfer pipette

1

25

IFU Leaflet

1

1

Material Required But Not Provided:

Timer 

Composition:

 

Conjugate Pad

Monoclonal Anti-SARS-CoV-2 antigen conjugated on the membrane.

G Line

Anti-human IgG

M Line

Anti-human IgM

C Line

Goat anti-rabbit IgG

Sample Buffer

0.01M PBS; PH 7.4

 

STORAGE AND STABILITY

1. Store the detector buffer at 2-30˚C.
2. Store the Healfo SARS-CoV-2 IgG/IgM Cassette Rapid Test at 2-30˚C; It can be stable until the expiration date.
3. If stored at 2-8˚C, ensure that the test device is brought to 15-30˚C before opening.
4. Do not freeze the kit or store the kit over 30˚C.

SPECIMEN COLLECTION AND PREPARATION

Consider any materials of human origin as infectious and handle using standard biosafety procedures.

Collection:

Serum or Plasma or Whole Blood

No special preparation or fasting of the patient is necessary. Serum or plasma derived from citrate or EDTA (ethylenediaminetetraacetate) as anticoagulants may be used.

Drops of whole blood can be obtained by venipuncture. Do not use hemolyzed blood for testing. Currently, no experimental data exists to support the use finger stick specimens.

Storage:

Serum or plasma specimens should be tested as soon as possible after collection. If specimens are not tested immediately, store at 2-8°C for up to 7 days. For long-term storage, specimens should be frozen at –20°C or colder. Specimens repeatedly frozen and thawed more than five (5) times or those containing particulate matter may give erroneous results.

Whole blood specimens should be stored at 2-8˚C if not tested immediately. The specimens must be tested within 24 hours after collection.

TEST PROCEDURE

Step 1: For fresh samples, begin with Step 2. For frozen samples, bring the specimens and test components to room temperature, and mix the specimen well once thawed.

Step 2: When ready to test, open the pouch at the notch and remove the test device. Place the test device on a clean, flat surface.

Step 3: Label the device with specimen ID number.

Step 4: Using a transfer pipette, transfer serum, plasma or whole blood, careful not to exceed the specimen well. The volume of the specimen is around 10μL. For better precision, transfer specimen by a pipette capable of delivering 10μL of volume.

Holding the transfer pipette vertically, dispense 10µL of the specimen into the center of the sample well (S well) making sure that there are no air bubbles.

Then, add 2 drops of Sample Diluent immediately into the sample well (S well).

Step 5: Set up a timer.

img3

Step 6: Read the results in 15-20 minutes.

Don’t read results after 20 minutes. To avoid confusion, discard the test device after interpreting the result. 

INTERPRETATION OF RESULTS

1.Valid Assay

img4

1.1 IgG positive & IgG weak Positive:If only the G Line and C Line are developed, this is result is IgG positive or reactive, in another word, that indicates the presence of IgG anti- SARS-CoV-2 virus, consistent with a recent or previous infection.

img5

1.2 IgM positive & IgM weak Positive:If only the M Line and C Line are developed, this is result is IgM positive or reactive, in another word, that indicates the presence of IgM anti- SARS-CoV-2 virus, consistent with an acute or recent SARS-CoV-2 virus infection.

img6

1.3 IgG & IgM positive and IgG & IgM weak Positive:If all of G, M Lines and C Line are developed, this is result is IgG and IgM positive or reactive, in another word, that indicates the presence of IgG and IgM anti- SARS-CoV-2 virus, suggesting current or recent SARS-CoV-2 virus infection.

img7

1.4 Negative :If only the C band is present, the absence of any burgundy color in the both test bands (G and M) indicates that no anti-SARS-CoV-2 virus antibodies are detected. The result is negative or non-reactive.

2. Invalid Assay:

img8

If the C Line does not develop, the assay is invalid regardless of color development of the G or M Lines as indicated below. Repeat the assay with a new device.

PERFORMANCE CHARACTERISTICS

Clinical Performance:

1.Study of: Testing of RT-PCR positive clinical specimens

 Ninety-eight (98) positive serum or plasma samples collected from individuals who tested positive with a RT-PCR method for SARS-CoV-2 infection and were quarantined in a makeshift hospital were used in this study. These patients, at the time of sample collection, exhibited mild or no clinical symptoms. These samples, along with 180 negative serum or plasma samples collected prior to September 2019, were tested together with the SARS-CoV-2 IgG/IgM Rapid Test. Of the 98 positive samples, ninety-one (91) were tested positive with IgG or IgM or both. Of the 180 negative samples, one hundred seventy-five (175) were tested negative.

Another 30 samples were collected from hospitalized individuals who were clinically confirmed positive for SARS-CoV-2 infection and exhibited severe symptoms. These samples, along with 70 negative serum or plasma samples collected prior to September 2019, were tested together with the SARS-CoV-2 IgG/IgM Rapid Test. Of the 30 positive samples, twenty-nine (29) were tested positive with IgG or IgM or both. Of the 70 negative samples, sixty-five (65) tested negative. The day of collection relative to the onset of illness was unknown.

Healfo SARS-CoV-2 IgG/IgM Rapid Test

Comparator

Subtotal

Pos

Neg

62

Positive

IgG+/IgM+

62

0

47

IgG-/IgM+

43

4

21

Negative

IgG+/IgM-

15

6

248

IgG-/IgM-

8

240

378

Subtotal

128

250

62

Taken together:

Positive Percent Agreement (PPA)= 120/128 (93.8%), 95% CI: 88.2% to 96.8%

Negative Percent Agreement (NPA)= 240/250 (96.0%),95% CI: 92.8% to 97.8%

2.Study of: Venous Whole blood specimens spiked with positive samples

Fifty (50) negative whole blood samples were spiked with positive serum at 1:100. Another fifty (50) whole blood specimens were spiked with negative serum at the same dilution. These 100 specimens were coded and tested with the SARS-CoV-2 IgG/IgM Rapid Test. All spiked samples were correctly identified by the test except for one of the negative samples, which was tested positive with the test. Thus, there was a 99% concordance rate with expected results when venous whole blood specimens are used.

Assay Cross Reactivity:

A low titer sample was diluted 1:100 to a serum or plasma sample containing antibodies reactive to one of following pathogens were tested along with unspiked samples in duplicate. No false positivity or false negativity was found: Human coronavirus(collected before Oct 2019)、HBV、HCV、HIV-1、HIV-2、Adenovirus、Human Metapneumovirus (hMPV)、Parainfluenza virus 1-4、Influenza A、Influenza B、Enterovirus 71、Respiratory syncytial virus、Rhinovirus、Chlamydia pneumoniae、Streptococcus pneumoniae、 Mycobacterium tuberculosis、Mycoplasma pneumoniae、EB Virus.

Potentially Interference Substances:

A low titer positive serum sample or negative serum sample was spiked with one of the following substances to specified concentrations and tested in duplicate. No false positivity or false negativity was found: Hemoglobin 10 mg/mL、Bilirubin Conjugated 0.4 mg/mL、Bilirubin Unconjugated 0.4 mg/mL、 Triglycerides 15 mg/mL、Cholesterol 4 mg/mL、Human Anti-mouse Antibody 800 ng/mL、 Rheumatoid Factor 2000 IU/mL、Human Serum Albumin 60 mg/mL、Histamine hydrochloride 4 mg/L 、α-IFN 200 mg/L、Zanamivir 1 mg/L、Oseltamivir carboxylate 1 mg/L、Abidol 40 mg/L、 Levofloxacin 200 mg/L、Ceftriaxone 400 mg/L、Meropenem 200 mg/L、Tobramycin 10 mg/L、 Ribavirin 40 mg/L、Human IgG 8 mg/mL、Human IgM 0.4 mg/mL.

QUALITY CONTROL

1. Internal Control: This test contains a built-in control to satisfies the quality control requirements. The C Line develops after addition of the specimen and sample diluent. If the C Line does not develop, the test is invalid, indicating that the test should be repeated.

2. Positive and Negative Control: Positive and negative controls should be tested to ensure the proper performance of the assay, particularly under the following circumstances: 1) new kits (new lot or new shipment); 2) new user; 3) new test environment (e.g., natural light vs. artificial light).; 4) abnormal storge environment (outside of 2-30°C); 5) abnormal working environment (outside of 15-30°C); 6) To investigate the cause of repeated invalid results; 

WARNINGS AND PRECAUTIONS

Inadequate adherence to package insert instructions may result in erroneous results.

1.Caution: Handle all relative biological materials as though capable of transmitting infectious agents.

2.Specimens should not be transported under extreme adverse temperature conditions.

3.Kits should not be used past their expiration dates.

4.All clinical specimens and materials used to collect these specimens should be considered potentially infectious and handled accordingly.

5.The assay should be performed at 15oC to 30oC.

6.Do not smoke, eat, or drink in areas where specimens or kit regents are handled.

7.Use disposable gloves and handle all materials used in the test (including samples, and controls) cautiously as though capable of transmitting infectious agents.

8.Dispose of all materials that have come into contact with specimens and reagents in accordance with local, state, and federal regulations.

LIMITATIONS OF PROCEDURE

1.Serum or plasma derived from sodium citrate, sodium heparin, or EDTA (ethylenediaminetetraacetate) as anticoagulants or whole blood may be used with this assay. Using other types of samples may not yield accurate results.

2.A test result that is INVALID should not be reported and the sample(s) should be retested.

3.The Healfo SARS-CoV-2 IgG/IgM Rapid Test is a qualitative assay, and the intensity of the test line does not necessarily correlate to SARS-CoV-2 antibody titer in the specimen.

4.A negative or non-reactive result can occur if the quantity of antibodies for the SARS-CoV-2 virus present in the specimen is below the detection limit of the assay, or the virus has undergone minor amino acid mutation(s) in the epitope recognized by the antibody utilized in the test.

5.If symptoms persist and the result from the SARS-CoV-2 Rapid Test is negative or non-reactive, it is recommended to re-sample the patient a few days later or test with an alternative test device.

6.The results obtained with this test should only be interpreted in conjunction with clinical findings, and the results from other laboratory tests and evaluations.

7.This test should not be used for screening of donated blood

8.Negative results do not rule out SARS-CoV-2 infection, particularly in those who have been in contact with the virus. Follow-up testing with a molecular diagnostic should be considered to rule out infection in these individuals.

9.Results from antibody testing should not be used as the sole basis to diagnose or exclude SARS-CoV-2 infection or to inform infection status.

10.Positive results may be due to past or present infection with non-SARS-CoV-2 coronavirus strains, such as coronavirus HKU1, NL63, OC43, or 229E.  

INdEX of CE SYMBOLS

 

 

 

MANUFACTURED BY

生产厂家Healvet Medtech GZ Ltd.

Add: B201, Building 2, No.6 Xinrui Road, Hi-Tech Industrial Development Zone, Guangzhou, P.R.China

Tel: +86-020-61855600 / +86-13828472609

Email: Healvet@healfo.com

欧盟代理CMC Medical Devices & Drugs S.L.

C/ Horacio Lengo N18, CP29006, Málaga-Spain

info@cmcmedicaldevices.com

DOC Version No.: HF0013C-20210309EN